Fluke T5-600 Voltage, Continuity and Current Tester
Specifications |
Jaw opening |
12.9 mm (0.5 in) |
Max wire size |
1/0 THHN Cable |
Current range ac rms |
0 to 100.0 A |
Accuracy ac current (50/60 Hz) |
3 % ± 3 counts |
AC Response |
Averaging |
Voltage range ac |
0 to 600 V |
Accuracy ac voltage |
1.5 % ± 2 counts |
Voltage range dc |
0 to 600 V |
Accuracy dc voltage |
1 % ± 1 count |
Resistance range |
0 to 1000 Ω |
Continuity beeper threshold |
On at <25 Ω, off at >400 Ω |
Volts indicator LED threshold |
Guaranteed on by 30 V ac |
Environmental Specifications |
Operating Temperature |
-10 °C to 50 °C |
Storage Temperature |
-30 °C to 60 °C |
Humidity (Without Condensation) |
0 % to 95 % (5 °C to 30 °C); 0 % to 75 % (30 °C to 40 °C);
0 % to 45 % (40 °C to 50 °C) |
Dust/water resistance |
Yes |
Operating Altitude |
2000 meters max |
Temperature Coefficient |
0.1 x (specified accuracy)/°C (<18 °C or >28 °C) |
Safety Specifications |
Overvoltage category |
CAT III 600 V ac/dc (EN61010-1) |
Agency Approvals |
UL, CSA, and VDE |
Maximum voltage |
Between any terminal and earth ground |
Mechanical and General Specifications |
Size |
30.5 mm x 51 mm x 203 mm |
Weight |
300g |
Warranty |
Two-years |
Battery Life |
200 hour zinc chloride, 400 hour alkaline |
Battery type |
2 x AA |
Low battery indicator |
Yes |
Current sensor opening |
12.9 mm |
Calibration cycle |
1 year |
Test leads |
Type: |
Heavy duty, flexible leads rated for 1000V use.
Field replaceable leads terminate in male shrouded banana plugs |
Probes: |
One red, one black.
Detachable Slim-Reach™ probe tips |
Styles: |
TP1 Slim Reach™ Test Probes |
Safety Specifications |
Safety |
CAT IV 600 V, CAT III, 1000 V |
Maximum voltage |
1000 V AC rms from earth to ground, in compliance with EN61010 |
Fluke 62 MAX+ Infrared Thermometer
Specifications |
Temperature Range |
-30 °C to 650 °C (-22 °F to 1202 °F) |
Accuracy |
±1.0 °C or ±1.0% of reading, whichever is greater
-10 °C to 0 °C: ±2.0
-30 °C to -10 °C: ±3.0 |
Response Time (95%) |
Fluke 62 MAX+ |
<300 ms (95% of reading) |
Spectral Response |
8 to 14 microns |
Emissivity |
0.10 to 1.00 |
Optical Resolution |
12:1 (calculated at 90% energy) |
Display Resolution |
0.1 °C (0.2 °F) |
Repeatability of Readings |
±0.5% of reading or <±0.5 °C (1 °F), whichever is greater |
Power |
AA battery |
Battery Life |
8 hours with laser and backlight on |
Physical Specifications |
Weight |
255 g (8.99 oz) |
Size |
(175 x 85 x 75) mm (6.88 x 3.34 x 2.95) inches |
Operating Temperature |
0 °C to 50 °C (32 °F to 122 °F) |
Storage Temperature |
-20 °C to 60 °C (-4 °F to 140 °F), (without battery) |
Operating Humidity |
10 % to 90 % RH non-condensing at 30 °C (86 °F) |
Operating Altitude |
2000 meters above mean sea level |
Storage Altitude |
12,000 meters above mean sea level |
IP Rating |
IP 54 per IEC 60529 |
Drop Test |
3 meters |
Vibration and Shock |
IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms |
EMC |
EN 61326-1:2006 EN 61326-2:2006 |
Standards and Agency Approval |
Compliance |
EN/IEC 61010-1: 2001 |
Laser Safety |
FDA and EN 60825-1 Class II |
|
|
|
|